MEASure:POWer[:REAL]?
This query returns the true power delivered to the unit under test by the AC source. Note that
the output relay must be closed to obtain current flow.
Query Syntax MEASure:POWer[:REAL]?
Parameters None
Examples MEAS:POW?
Returned Parameters <NR2>
Related Commands MEAS:VOLT? MEAS:CURR?
MEASure:POWer:APParent?
This query returns the apparent power delivered to the unit under test by the AC source.
Note that the output relay must be closed to obtain current flow.
Query Syntax MEASure:POWer:APParent?
Parameters None
Examples MEAS:POW:APP?
Returned Parameters <NR2>
Related Commands MEAS:POW?
MEASure:POWer:PFACtor?
This query returns the power factor. The power factor is the ratio between real power and
apparent power. This value is always between 0.00 and 1.00. If the current measured is too
low to determine the power factor, a 1.00 value is returned.
Query Syntax MEASure:POWer:PFACtor?
Parameters None
Examples MEAS:POW:PFAC?
Returned Parameters <NR2>
Related Commands MEAS:POW? MEAS:POW:APP?