AMETEK Programmable Power MX / RS Series SCPI Programming Manual
54 MX / RS Series
4.5 Array Measurement Subsystem [3Pi Controller Only]
This command subsystem lets you retrieve arrays containing measurement data. Only
current and voltage measurements are stored in an array. Two measurement commands
are available: MEASure and FETCh. A MEASure command triggers the acquisition of new
data before returning the readings from the array. A FETCh command returns previously
acquired data from the array.
Individual outputs of a three-phase source are specified by the setting of
INSTrument:NSELect.
Subsystem Syntax
MEASure | FETCh
:ARRay
:CURRent
[:DC]? Returns the digitized instantaneous current
:HARMonic
[:AMPLitude]? Returns amplitudes of the first 50 harmonics
:PHASe? Returns phase angles of the first 50 harmonics
:MODE Selects waveform data transfer format
:VOLTage
[:DC]? Returns the digitized instantaneous voltage
:HARMonic
[:AMPLitude]? Returns amplitudes of the first 50 harmonics
:PHASe? Returns phase angles of the first 50 harmonics